ZY CleanroomTIANJIN · SINCE 2002 Get a budget range

Buyer guide

VHP pass box: how the cycle works and what to specify

A VHP pass box is a sealed transfer chamber that decontaminates the outside of items with vapour-phase hydrogen peroxide (VHP) before the inner door opens. Where it is used, EU GMP Annex 1 (para 4.36) asks that the effectiveness of the agent and the dispersion system be understood and validated. The label "VHP" on a box does not prove that its cycle works for your load. That proof comes from cycle development and biological indicator tests with the items, packaging and loading pattern you actually use.

By ZY Cleanroom · Updated · Sources at the end of the page

What a VHP pass box is for

A static pass box moves items through a wall. A dynamic pass box adds a fan and HEPA filter that flush the chamber with filtered air. A VHP pass box adds a decontamination step for the surfaces of the items.

Annex 1 sets the rules that apply to any pass-through hatch:

RuleAnnex 1 paragraph
Entry and exit doors are not opened at the same time4.13
Interlocking system for airlocks leading to grade A and B; at least a visual or audible warning for grade C and D4.13
Hatches protect the higher-grade side, for example by flushing with an active filtered air supply4.12
Only items on an approved list, assessed during validation of the transfer process, go into grade A or B4.12
Items moving from a lower grade to a higher grade are cleaned and disinfected in line with the risk and the contamination control strategy (CCS)4.12
Fumigation or vapour disinfection (for example VHP) is understood and validated, agent and dispersion system4.36

None of these paragraphs says a VHP box is needed for every transfer. The owner's CCS decides which transfers need a decontamination cycle. Typical reasons are items that cannot be wiped well, a high transfer frequency into grade A or B, or a CCS that asks for a sporicidal step at the boundary.

VHP pass box: how the cycle works and what to specify

How a VHP cycle works

A VHP cycle usually runs in four stages. Ask the supplier to name the stages of the offered model and the set points for each one.

StageWhat happensWhat to ask for
Preparation (conditioning)The chamber is closed and its air is brought to the starting condition the cycle needs, often a set humidityStart condition and how it is measured
GassingHydrogen peroxide vapour is injected and spread through the chamberInjection rate, the concentration reached and how it is measured
DwellThe concentration is held for a set timeHold time and the alarm if the concentration falls
AerationThe vapour is removed until the residual level is low enough to open the doorResidual level that releases the inner door, and how it is measured

Two points matter more than the numbers:

  • Concentration alone is not proof. A cycle can reach its set concentration and still miss a shadowed surface. Biological indicators show whether the cycle killed the challenge organism at each position.
  • Cleaning comes first. For isolators and RABS, Annex 1 (para 4.22) says the cleaning before bio-decontamination is essential, because residues can inhibit the decontamination. The same logic applies to items going into a VHP box: wipe them first, then run the cycle.

Biological indicators and the kill target

Annex 1 defines a biological indicator (BI) as a population of microorganisms on a carrier, placed in a load or room location to show whether a sterilisation or disinfection cycle works. The challenge organism is chosen for its resistance to that process.

For hydrogen peroxide cycles, a common target is a 6-log reduction of a resistant spore BI at the hardest-to-reach positions. If the BI's D-value is known, the time to reach that reduction is six times the D-value. This is a usual design target, not a figure Annex 1 sets. The owner's quality team sets the acceptance criterion in the protocol.

Place BIs where the vapour reaches last: under the item, in corners, at the door seal and inside packaging folds. Run the tests with the real load, because the packaging and the loading pattern change how the vapour spreads.

Safety: residual hydrogen peroxide

The inner door should not release until the residual hydrogen peroxide in the chamber is below the level the owner sets. For reference, the US OSHA permissible exposure limit for hydrogen peroxide is 1 ppm as an 8-hour time-weighted average. Local limits apply on your site; check them with your safety officer.

Ask the supplier how the residual is measured, where the sensor sits, how often it is calibrated, and what the box does if aeration does not reach the release level.

What to put in the URS

FieldExample entry
Mark and locationPB-03, grade C corridor to grade B room
Transfer listApproved items, each with packaging and size
Largest load, W × H × DWith packaging or tray, mm
Usable chamber and clear door openingTwo separate numbers, mm
FunctionVHP decontamination cycle, plus filtered-air flushing if the CCS asks for it
InterlockDoors never open together; inner door locked until the cycle and aeration are complete
Cycle dataStages, set points, hold time, release level for residual
Hydrogen peroxide supplyBuilt-in generator or connection to a site generator; agent concentration
UtilitiesVoltage and frequency, compressed air if needed, exhaust route if any
Chamber material and finishState the grade and finish you need and the agents it must tolerate (Annex 1 para 4.7)
DocumentsDrawings, cycle description, operating and maintenance instructions, sensor calibration records
TestsInterlock check, leak tightness of the chamber, sensor checks; cycle development and BI tests for your load

Keep chamber size, clear opening and external size as three separate numbers. Check the installation opening against the wall thickness and the loading height on both sides.

Who does what

The supplier delivers the box, its cycle description, documents and the factory tests agreed in the quotation. The owner's quality team writes and approves the qualification protocol, sets the acceptance criteria and approves the result. For the DQ, IQ, OQ and PQ chain, see the pass box qualification guide.

VHP pass boxes are part of the ZY Cleanroom product range. We supply them with your panels and doors in one quotation, so sizes and openings match, and each unit is inspected before shipping. Cycle validation for your load is a separate scope, defined in writing.

Questions

Common questions

What is a VHP pass box?

A sealed transfer chamber that runs a vapour-phase hydrogen peroxide cycle on the surfaces of items before the inner door opens. Its doors are interlocked so that they never open together.

How does a VHP pass box work?

The items are wiped and loaded, the outer door closes, and the box runs its cycle: preparation, gassing, dwell and aeration. The inner door unlocks only when the cycle has finished and the residual hydrogen peroxide is below the release level.

Is a VHP pass box the same as a sterilising pass box?

No. VHP decontaminates exposed surfaces. It does not sterilise the inside of packages or products. Items for grade A that must be sterile are sterilised by a validated method and enter in sealed packaging (Annex 1 para 8.47).

Does a VHP pass box need a HEPA filter?

It depends on the model and on your CCS. Annex 1 para 4.12 gives flushing with an active filtered air supply as an example of how a hatch protects the higher grade. If you need that flushing as well as the VHP cycle, write both functions in the URS.

How is a VHP pass box validated?

Develop the cycle with your real load, prove distribution with biological indicators at the worst positions, and confirm that aeration reaches the release level. Annex 1 para 4.36 asks that the agent and the dispersion system be understood and validated. The owner's quality team approves the protocol and the result.

Can a UV lamp replace the VHP cycle?

No. Annex 1 para 8.71 says ultraviolet irradiation is not an acceptable method of sterilisation. If a box has a UV lamp, do not count it as the transfer control.

What should I check when choosing a VHP pass box manufacturer or supplier?

Ask for the cycle description with stages and set points, how concentration and residual are measured, the interlock logic, the chamber material and finish, the drawings and manuals, and which factory tests come with the unit. Then ask what cycle development and biological indicator testing for your load is included and what is a separate scope. Compare every offer on the same URS.

How long does a VHP cycle take?

It depends on the chamber size, the load and the aeration method, so this guide gives no figure. Ask for the cycle time with your load, including aeration, and check it against how many transfers you need per shift.

Related guides

Sources

Project enquiry

Send what you have

A room list, a floor plan or only the floor area is enough to start. After you submit, send drawings by email or WhatsApp.

We reply by email within one working day. We can sign an NDA before you send drawings.

WhatsApp +44 7962 561184 is our UK number for export enquiries. The factory is in Tianjin, China.

Fastest reply: message us on WhatsApp. Or use the form, and add your WhatsApp number so we can reply there.

Email: [email protected]
WhatsApp: +44 7962 561184

Step 1 of 2

Chat on WhatsAppGet a budget range