Pass box and air shower interlocks: how they work and what to specify

An interlock stops the two doors of a pass box or air shower from being open at the same time. EU GMP Annex 1 (para 4.13) asks for an interlocking system where the transfer leads to grade A or B, and at least a visual or audible warning where it leads to grade C or D. Specify more than "interlocked": state the time delay, what each door does in a power cut, how the emergency release works, and which tests prove it.

By ZY Cleanroom · Updated · Sources at the end of the page

What the texts require

PointTextSource
Doors open togetherEntry and exit doors of pass-through hatches and airlocks should not be opened simultaneouslyEU GMP Annex 1 (2022) para 4.13
Into grade A or BAn interlocking system should be usedAnnex 1 para 4.13
Into grade C or DA visual and/or audible warning system, as a minimumAnnex 1 para 4.13
Time delaySet a delay between closing one interlocked door and opening the other where it is needed to keep areas apartAnnex 1 para 4.13
Pharmacy pass-throughBoth doors never open at the same time, and the doors should interlockUSP <797> 4.2.1
DefinitionA pass box has a door on each side that may not be opened at the same timeGB 50073-2013, term 2.0.22

Annex 1 does not mention air showers. If you fit one, treat its two doors as an interlocked pair, the same as an airlock.

Pass box and air shower interlocks: how they work and what to specify

How an interlock works

Two arrangements are common. Ask which one the offered model uses.

Mechanical interlockElectronic interlock
How it worksA linkage in the frame blocks one handle while the other door is openA sensor on each door and a lock on each door, run by a controller
PowerNoneNeeds a supply; behaviour in a power cut must be defined
Time delayNot possibleSet in the controller
IndicatorsNone, or a simple flagLamps or a display on both sides; buzzer
Typical useStatic pass boxes between rooms of similar riskDynamic and VHP pass boxes, air showers, transfers into grade A or B

A mechanical interlock cannot hold a door shut while a flush or a decontamination cycle runs. If the inner door must wait for a cycle to finish, the interlock needs a controller.

Time delay: tie it to the process

Annex 1 asks for a delay only where it is needed. Set its length from what happens inside the box or shower:

  • Dynamic pass box: the inner door waits until the chamber has been flushed with filtered air for the time set in your procedure. Annex 1 para 4.12 gives flushing as an example of how a hatch protects the higher grade.
  • VHP pass box: the inner door waits until the cycle and aeration are complete and the residual is below the release level. See the VHP pass box guide.
  • UV lamp: if a box has a lamp, the lamp should switch off when a door opens. Do not count the lamp as the transfer control; Annex 1 para 8.71 says ultraviolet irradiation is not an acceptable method of sterilisation.
  • Air shower: the inner door unlocks when the blow cycle ends.

Write the delay in seconds in the data sheet, and say who may change it.

Power cuts, emergency release and fire alarms

These three points are often missing from a specification, and each one needs an answer before the order.

QuestionWhy it mattersWhat to write
Power cut: do the doors lock or release?Locked doors keep the rooms apart but can trap a person in an air showerThe state of each door with no power
Emergency releasePeople must be able to leave an air showerType and position of the release on the inside; whether it raises an alarm
Fire alarm signalEscape routes may need to openWhether the fire alarm releases the doors, agreed with the local fire authority
Bypass key or switchMaintenance needs both doors openWho holds it and how its use is recorded

A pass box is a hatch, not an escape route, so its doors can stay locked in a power cut. An air shower is a personnel door, so local escape rules apply to it. Agree the release behaviour with your safety officer and fire authority.

The data sheet line

FieldExample entry
MarkPB-02 / AS-01
From / toGrade C to grade B
InterlockElectronic; doors never open together
DelayInner door released 60 s after outer door closes (example value; set by your procedure)
IndicatorsDoor status lamp on both sides; buzzer if a door is held open
Power cutPass box: both doors stay locked. Air shower: both doors release
Emergency releaseAir shower: inside push button, raises an alarm
Fire alarmAir shower doors release on fire alarm signal, if the fire authority requires it
BypassKey switch, held by maintenance, use logged
TestsInterlock both ways, delay, indicators, alarm, power-cut behaviour, emergency release

Testing the interlock

During OQ, test what the data sheet says the interlock does, not just that one door locks:

  • Open door A: door B stays locked. Then the reverse.
  • Close door A: door B stays locked until the delay ends.
  • Hold a door open: the alarm sounds after the set time.
  • Cut the power: each door behaves as written.
  • Press the emergency release: the door opens and the alarm is raised.

The pass box qualification guide lists these with the airflow, filter and pressure tests. We provide IQ and OQ documents for the units we supply; your quality team approves the protocol and the result.

One quotation for the whole entry

Pass boxes and air showers are part of our cleanroom product range. We supply them with your panels and doors in one quotation, so the wall openings, wall thickness and interlock schedule are planned together, and each unit is inspected before shipping. Send your room list or door and pass box schedule: Chat on WhatsApp (+44 7962 561184), or use the form below.

Questions

Common questions

What is a pass box interlock?

A device that stops the two doors of a pass box opening at the same time. While one door is open, the other stays locked.

Is an interlock required on every pass box?

Annex 1 para 4.13 asks for an interlocking system into grade A and B, and at least a visual or audible warning into grade C and D. USP <797> says pharmacy pass-through doors should interlock. In every case, the two doors should not be open together.

Mechanical or electronic interlock?

A mechanical interlock needs no power but cannot add a time delay or wait for a cycle. Choose electronic for dynamic and VHP pass boxes, air showers and transfers into grade A or B.

What does a pass box controller do?

It reads the door sensors, drives the door locks, runs the delay and any fan or cycle, and shows the door status on both sides.

What happens to an air shower in a power cut?

Whatever the data sheet says. Because people stand inside, the usual choice is that the doors release, with an emergency release on the inside. Agree it with your safety officer.

How long should the interlock delay be?

No text sets a figure. Set it from the flush, blow or decontamination time in your procedure, and confirm it in OQ.

Related guides

Sources

Project enquiry

Send what you have

A room list, a floor plan or only the floor area is enough to start. After you submit, send drawings by email or WhatsApp.

We reply by email within one working day. We can sign an NDA before you send drawings.

WhatsApp +44 7962 561184 is our UK number for export enquiries. The factory is in Tianjin, China.

Fastest reply: message us on WhatsApp. Or use the form, and add your WhatsApp number so we can reply there.

Email: [email protected]
WhatsApp: +44 7962 561184

Step 1 of 2

Chat on WhatsAppGet a budget range